Features of the instrument:
- Designed to be used for CO monitoring, where very fast measurements and low detection limits are required.
- Fully automated operation using 166 MHz PENTIUM CPU with internal data storage on hard-drive or floppy
- Automated calibration using external CO-standards
- Linearity range (pulse counting): 0 to >100 ppm
- Detection Limit:
- < 1.5 ppb @ 1s
- < 0.8 ppb @ 10s
- Rise- and fall time (10 to 90 %) < 0.1 s
- Low weight (24 Volt DC-air craft version available) < 22 kg
- Rugged and simple to use.
- RS232- and LAN-interface to external PC
|