Instrument Specifications
The instrument has multiple, free programmable measuring ranges. The calibration can be defined by user within the ranges given below. Liquid phase measurements provides calibration by liquid peroxide standard for example:
: 0 to 30.0 µg/l : 0 to 300.0 mg/l : 0 to 3.0 mg/l*
* Higher concentrations on request.
|
Detection limit:
|
< 0.07 µg/l**
|
|
|
equivalent to: 2 * 10-9 M
|
|
|
** only when calibrated for 30.0 µg/l
|
|
Noise (each channel)
|
< 2% at full scale
|
|
Time constant
|
40 sec (10-90%)
|
|
|
delay time 240 sec
|
|
Interferences from
|
|
|
O3
|
+30 ppt H2O2/100 ppb
|
|
NO
|
+12 ppt H2O2/100 ppb
|
|
|
|
|
No detectable interference from:
SO2, PAN, NO2, Glyoxal, Isobutane, Isobutylene, 1-Butane, Formaldehyde, Benzene, Toluene, Methanol, Acetone, Methylamine, Dimethylamine, n-Butane, Cis-2-Butene, Trans-2-Butene, and Ions such as Iodide, Chloride, Nitrate, Bromide, Phosphate, Benzoate
|
|
Zeroing:
|
Using Water free of peroxides
|
|
Calibration:
|
Using liquid H2O2-standards
|
|
Signal Output
|
|
|
Analog:
|
0-5 Volts FS. calibration sets automatically output to 4.0 Volt of maximum value of chosen range.
|
|
Digital:
|
via RS 232 interface
|
|
Operating Temp.:
|
20 to 40 degrees C
|
|
|
5 to 95% rel. humid
|
|
Physical Dim.:
|
Length 49,6 cm
|
|
|
Weight 48,8 cm
|
|
|
Height 13,0 cm
|
|
|
Weight 22,5 Kg
|
|
Power Requirem.:
|
120/240VAC, 80 Watts
|
|
|
(18 to 36 V DC on request).
|
|
Consumables
|
Destilled Water
|
|
|
Solution AL1 and AL2, NaOH, HCl,
|
|
|
H2O2-standard solution for calibration.
|
|
Maint. Interval:
|
1 Month
|
|
|
|
|